Table 3 (support for IEC 60601-1-2:2007)
Guidance and manufacturer's declaration - electromagnetic immunity
The DRYPRO 832 is intended for use in the electromagnetic environment specified below. The customer or the
user or the DRYPRO 832 should assure that it is used in such an environment.
Immunity test
Conducted RF
3Vemf 150 kHz
IEC 61000-4-6
to 80 MHz
Radiated RF
3V/m 80 MHz to
IEC 61000-4-3
2.5 GHz
[NOTE] At 80 MHz and 800 MHz, the higher frequency range applies.
[NOTE] These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theo-
retically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the location in which the DRY-
PRO 832 is used exceeds the applicable RF compliance level above, the DRYPRO 832 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re- orienting or relocation the DRYPRO 832.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3]V/m.
IEC 60601 test
Compliance
level
level
[3]V
[3]V/m
21
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the DRYPRO
832, including cables, than the recommended separa-
tion distance calculated from the equation applicable
to the frequency of the transmitter.Recommended
separation distance
d=[1.2] √ P
d=[1.2] √ P80 MHz to 800 MHz
d=[2.3] √ P800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m)
Field strengths from fixed RF transmitters, as de-
termined by an electromagnetic site survey,
be less than the compliance level in each frequency
range.