Konica Minolta DRYPRO 832 Operation Manual page 68

Laser imager
Hide thumbs Also See for DRYPRO 832:
Table of Contents

Advertisement

Operation of Test Print
Density Check
Check the density value according to the items to be checked shown below.
Item
Category
Number
When Checking from the Message Display Window on the Operation Panel
(11)
Density
(12)
Check the readout given by the densitometer built into the main unit from the operation panel of the
main unit.
When Checking from the Message Display Window on the Operation Panel
1
Check that "Dens. Error <Tray No.>" is not displayed on the DRYPRO 832
operation panel. (Item to be checked:
When QC patterns are printed from the standard supply tray (tray 1)
2
Press the
The QC Results View of the standard supply tray (tray 1) is displayed.
When QC patterns are printed from Lispl-832 (tray2)
2
Press the
The QC Results View of the Lispl-832 (tray 2) is displayed.
3
Check that "NG" is not displayed in "1" to "4" of the "QC Result" and the
correct density value (three digits) is displayed. (Item to be checked: )
After the expiration of 30 seconds without external input, the screen is restored to
the Current Status View.
After printing the QC patterns, "Dens. Error <Tray No.>" is not displayed on the DRYPRO 832
operation panel
After printing the QC pattern, "NG" is not displayed in "1" to "4" of the "QC Result" in the QC
Results View on the DRYPRO 832 operation panel, but the correct density value (three digits)
is displayed
[Reset/Cancel] button twice.
[Reset/Cancel] button three times.
Item to be Checked
)
68