Replacing the Battery
The battery can be replaced on site as needed, provided the following precautions and
instructions are strictly followed:
Precautions:
Use only approved MD.2 manufacturer supplied replacement battery. Contact your
MD.2 reseller for replacement battery.
The battery contains hazardous materials and must be disposed of properly. Your
MD.2 reseller can assist you if necessary.
Replacement instructions:
Removal of old battery
- The MD.2 does not need to be turned off when changing battery. If you
choose to leave it powered on, make sure the wall transformer is plugged in
properly before proceeding.
- Open side door.
- Push down retaining clamp, slide battery partially out (use plastic tab on
battery).
- Once battery cables are exposed, remove cables from battery terminals (2).
- Complete removal of battery.
Installation of New Battery
- Slide new battery partially into battery tray of MD.2, battery terminals
should be exposed and visible to the installer.
- Connect the RED wire to terminal labeled + . This terminal will be on the
right side of the battery.
- Connect the BROWN wire to the terminal labeled. This terminal will be on
the right side of the battery.
- Slide battery completely into its holding tray, making sure it slides past the
battery retaining clip.
WARNING
!
FAILURE TO CONNECT BATTERY CABLES PROPERLY CAN CAUSE DAMAGE TO THE
MD.2 AND THE BATTERY.
DO NOT SHORT CIRCUIT BATTERY TERMINALS, DAMAGE TO BATTERY/FIRE
HAZARD CONDITION WILL RESULT (this applies to both new and used batteries).
DO NOT CRUSH, INCINERATE OR OTHERWISE DISPOSE OF BATTERY IMPROPERLY.