Step 3. You will be removing the cover from the unit. If there is not enough headroom to do
this where the MD.2 is currently placed, move it to a suitable table or workspace.
Step 4. Remove the door: Open the door and swing it all the way back and then lift upwards.
The door will lift off its hinges. Place the door on a clean flat surface away from the
MD.2
Step 5. Remove the top cover: Gently pry off the round plug located on the top of the machine,
using the flat blade screwdriver, prying it from the door side of the plug. With the
Phillips screwdriver, unscrew the screw that holds the cover down (in the recess below
the plug that was just removed). When the screw is loose, lift the cover straight up, over
the medication cup cylinders of the MD.2. Place the cover on a clean flat surface away
from the MD.2
Step 6. With your fingers, slide all medication cups out of the medication cup cylinder
mechanism. Also free any cups stuck in the medication cup transporter (4" disk to the
left of the exposed speaker.
Step 7. If you are disassembling the unit because of a jammed cup, GENTLY check for free
movement of moving parts once cups are removed.

Re-assembling the Unit

Step 1. Place the cover back onto the unit. It is designed so it will only go on one way.
Step 2. While holding the cover in place, put the screw back into position and tighten the screw
only until it is hand tight. DO NOT OVERTIGHTEN as you could damage internal
components and fasteners.
Step 3. Snap the plug back into position on the top of the cover. Note that it is also designed to
only go on one way.
Step 4. Pick up the door and align the hinges with the door in the fully opened position. When
both hinge pins are aligned, the door will slip into place.
Step 5. If you successfully dislodged a cup from the unit, GENTLY move the medication cup
cylinder mechanism to make sure it moves properly. DO NOT FORCE the mechanism.
If the mechanism turns with a modest effort, you are ready to turn on the unit.. Follow
the directions in Section 4 – "Unpacking and Preparing Your MD.2 for Use" to re-
establish the power and telephone connections to the MD.2. You must reprogram (see
Section 5) the MD.2 and reload the medication cups.
If your unit still does not operate properly, please call our Support Center at (877) 472-9037.
They will arrange to ship a replacement unit to you right away.