Siemens MAMMOMAT Balance Troubleshooting Manual page 24

Table of Contents

Advertisement

• Calibration
During tube running measure Vdc between TP1 and GND (ground) , then adjust RV1 to have Vdc at
TP3 = 55% of Vdc at TP1.
Repeat same operation with TP2, RV2 and TP4.
Typical values of running voltages at TP1 and TP2 for IAE tube XM12 are :
Factory calibration voltages at TP1, TP2, TP3, TP4 can be found in the tube calibration report.
MAMMOMAT Balance
Troubleshooting Guide
VTP1
VTP2
SPB7-115.840.01
Page 24 of 62
Rev.02
10.04
Rotating anode starter
V(TP3) = 0.55xV(TP1)
V (TP4) = 0.55 x V (TP2)
IAE
1.2 V
2.5 V
Siemens AG
Medical Solutions