Dosemeter data

MAMMOMAT Balance
Troubleshooting Guide
You can define dosemeter instruments for use in fine tuning measure
readout. Just press the "Edit"-"Dosemeter data" menu.
SPB7-115.840.01
Page 47 of 62
Rev.02
10.04
Dose calculator
The "Dosemeters" window
opens, with edit fields all
editable. To define a new
instrument, press the "new"
menu. A "default" dosemeter is
inserted in the left list. Now you
can select it in the list (as with an
existing one) and type the
correct informations in
corresponding edit fields on the
right. To modify the response
curve for the instruments (the
default is a plateau curve) press
the "Edit calibration curve"
pushbutton.