Technical Menu - Siemens MAMMOMAT Balance Troubleshooting Manual

Table of Contents

Advertisement

ALARM
Code
When
Starting exposure
Why
X-ray tube assembly has reached maximum operating temperature.
Action
Wait for tube cooling.
Evaluate operating conditions from last 1300 exposure memory.
Check tube temperature sensor PCB 02-171.
Disable tube rotation brake.
ALARM
Code
When
Any time.
Why
Internal error having no meaning for the operator.
Action
Check Log File to find out what it is.
Date and time of error / fault can be helpful to locate it quickly

2) Technical menu

When
Any time.
Why
Broken Fuse or defective block has been located by CPU diagnostic functions
Action
Check Log File to find out when it occurred
MAMMOMAT Balance
Troubleshooting Guide
TUBE THERMAL LIMIT REACHED
TECHNICAL ERROR
SIEMENS AG
CPU 01-170
F2=OK
SW1 00000000
SW2 0000
INVERTER Vdc (in) = 545 V
Filament
92-073
Rotating Anode
92-074
FIRMWARE REV
BAL__ XX TUBE
Tube Housing kJ
n. ..............
../../.. ..................
SPB7-115.840.01
Page 18 of 62
Rev.02
10.04
HU
Refer to list
F4=OK
LV= OK
>> PASS <<
>> PASS <<
25°C
Troubleshooting