Memory String - Siemens MAMMOMAT Balance Troubleshooting Manual

Table of Contents

Advertisement

Troubleshooting
Now you can see, on the "Uploading statistical info" window, the data being uploaded. When the
mammo finishes to transmit data, it emits three quick tones to advise you. If you want to stop the
transmission push the two buttons Alarm reset and Optical density minus contemporarily. If you want
to delete all the 1300 events in the mammo, press the three buttons alarm reset, optical density plus
and optical density minus contemporarily.
DEATA Plus software creates a LOG file that can be opened clicking on View Stat button.
Log file contains string with error code and full description in the last line
Error log file can be selected between all those available, identified by time and date of Upload

Memory string

The following string is an example of data available for each exposure or fault.
----------------------------------------------------------------
>5 26.5kV 126.6mAs .30 Mo/Mos G 0P 27.2mGy\
>6 r0110 p076 i0472 d 0 c100 %100 o011 C1 \
Af
>7 kS206 r0110/0110/0110/0110 38°C
[12/03/03 15:20:02]\
----------------------------------------------------------------
It's referred to exposure in Zero Point mode with code Af ( no error in automatic mode)
MAMMOMAT Balance
SPB7-115.840.01
Page 21 of 62
Siemens AG
Troubleshooting Guide
Rev.02
10.04
Medical Solutions