Technical Data - Maico easyTymp Operating Manual

Hide thumbs Also See for easyTymp:
Table of Contents

Advertisement

easyTymp® Operating Manual

8 Technical Data

The easyTymp® is an active, diagnostic medical product according to the
class IIa of the EU medical directive 93/42/EEC
Standards:
Power, UE24WCP-
240100SPA
Mains voltages and fuses:
Power, easyTymp®
Power, Cradle
Operation
environment:
Transport:
Dimension and weight
Impedance Measuring System
Probe tone:
Air pressure:
Compliance:
Test types:
Indicators:
Safety:
EMC:
Impedance:
Consumption:
Fuses:
Fuses:
Temperature:
Rel. Humidity:
Storing/handling:
Temperature:
Rel. Humidity:
Dimension
Weight
Frequency:
Level:
Control:
Indicator:
Range:
Pressure limitation:
Pressure change rate:
Range:
Tympanometry
Graphical display
29
IEC 60601-1, Class II, Type B
IEC 60601-1-2
IEC 60645-5/ANSI S3.39, Type 1
0,6 A
100 – 240 VAC, 50 – 60 Hz
3 A (5 V)
1 A (24V)
15 – 35 C
30 – 90%
Temperatures below 0C and above 50C may
cause permanent damage on the instrument
and its accessories.
-20 – 50 C
10 – 95%
W x D x H: 80 x 360 x 60 mm
417 g
226 Hz, 1000 Hz
69 dB HL with AGC, assuring constant level at
different ear canal volumes.
Automatic.
Measured value is displayed on the graphical
display.
-300 to +200 daPa.
-750 daPa and +550 daPa.
Minimum, medium, maximum or automatic
with minimum speed at compliance peak.
Selectable in the set-up. Range is 50-350daPa
0.1 to 8.0 ml at 226 Hz probe tone (Ear volume:
0.1 to 8.0 ml) and 0.1 to 15 mmho at 1000 Hz
probe tone.
Automatic, where the start and stop pressure
can be user-programmed in the set-up function.
Compliance is indicated as ml and pressure as
daPa.
Stimulus level is indicated as dB Hearing Level.