5.2 Installation

1. Connect all the components according
to the figures shown on page 4 of this
manual. Use grounded power outlet.
Ensure the machine is on a flat
surface prior to operating.
2. Press the power switch "I" side and
switch on the instrument.

5.3 Settings

Speed Settings
1. Press the "
" button , the "rpm"
Mode
indicator light will be on.
2. Press the rotary knob, the number
shown on the Speed/Tilt display
window starts flashing, turn rotary
knob to increase or decrease the
speed value.
3. When the speed shown on Speed/Tilt
display window reaches the set value,
press the knob again within 5s to save
the value, and the display window will
stop flashing. The speed setting is
completed.
Tilt Angle Setting
1. Press "
" button and the "degree"
Mode
indicator light will be on.
2. Press the rotary knob, the number
shown on the Speed/Tilt display
window starts flashing, turn rotary
knob to increase or decrease the Tilt
value. The range of the tilt displayed
on the window is 1° to 24°.
9
Section 5 | Operation
3. When the angle shown on the
Speed/Tilt display window reaches the
set value, press the rotary knob again
within 5s to save the value, and the
display window will stop flashing. The
tilt angle setting is completed.
Run and Stop
Press "
Run
" button and the instrument
Stop
will start running .
Press "
" button again and the
Run
Stop
instrument will slow down until it stops.
The instrument will then be in standby
and the display window will show the set
values.
Finish Operation
After the operation is finished, press the
power switch at the back right side the
instrument and put it into the "O" state.
Unplug the instrument and store the
instrument according to the storage
guide.
Alarm System
E1: If running speed does not match the
setting speed for 3mins, there will be a
buzzer, and then E1 will be displayed on
the Speed display window .
When the instrument alarms, press any
key and the instrument goes into
standby mode.
Power Recovery
If the power supply is cut off suddenly
while the instrument is in operation, the
unit will automatically run at the
previously set parameter upon power
restoration. The display window will
flash. Press any button to stop the
flashing.
Mechanical Zero Correction
To ensure accurate and quality rocking,
the instrument will automatically run
mechanical zero correction in the
following situations:
1. First operation after the machine is
turned on.
2. First operation after the rocking speed
is reset.
3. After 8 hours of continuous operation.

5.4 Installation of Accessories

Only accessories that meet the
manufacturer's specifications shall be
used.
Installation of Double Tray
1. Place the Tube Rocker Stacking Bars on
the base tray .
2. Stack the Double tier tube platform on
the Bars.
3. Tighten screws using a screw driver.
4. Place the Rubber Mat on the tray.
Digital Tube Rocker