Invacare Platinum POC1-100B Service Manual page 26

Oxygen concentrator
Hide thumbs Also See for Platinum POC1-100B:
Table of Contents

Advertisement

Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
Battery
0xb20
LPA_batteryTooHot-
Too Hot
b
ToCharge
To Charge
(Running)
Battery
0xb20c
LPA_batteryTooCold-
Too Cold
ToCharge
To Charge
(Running)
Battery Hot
0x6205 LPA_batteryHot
(Running)
26
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Cannot Charge Allow
to Cool"
The display shows:
"Cannot Charge Allow
to Warm
The display shows:
"Battery Hot Allow to
Cool"
Possible Reason
Solution
1. The concentrator
1. Remove the hot
is running off
external power
(AC/DC) and
2. If alarm persists,
at least one
of the battery
temperatures is
greater than or
equal to 49° C.
1. The concentrator
1. Remove the cold
is running off
external power
(AC/DC) and
2. If alarm persists,
at least one
of the battery
temperatures is
less than or equal
to 8° C.
1. The concentrator is
1. Remove the hot
running off battery
power and at least
one of the battery
2. If alarm persists,
temperatures is
greater than or
equal to 55°C.
battery and allow
to cool.
contact Invacare
for service.
battery and allow
to warm.
contact Invacare
for service.
battery and allow
to cool.
contact Invacare
for service.
1187375-B

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents