Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
0x3201 LPA_noBreathDetected
No Breath
Detected
(Running)
Breath Rate
0x4202 LPA_breathRateTooHigh
Too High
(Running)
24
All manuals and user guides at all-guides.com
Display Message
The display shows:
"No Breath Detected
Check Cannula" (not
displayed if in "Warm
Up" or "Prolonged
Warm Up")
The display shows:
"Breath Rate High
Reduce Activity"
Possible Reason
Solution
1. The concentrator
1. Verify cannula is
has not detected
an inhalation for
more than 15
seconds. -or-
2. Restriction after
the conserver
valve.
2. Check the outlet
3. If alarm persists,
1. The user breath
1. Make sure breath
rate has exceeded
the capacity of
the concentrator
2. If alarm persists,
for more than 15
seconds.
connected, not
kinked, properly
positioned and
user is breathing
through his/her
nose.
filter/Replace filter.
contact Invacare
for service.
rate is lower than
40BPM.
contact Invacare
for service.