Invacare Platinum POC1-100B Service Manual page 36

Oxygen concentrator
Hide thumbs Also See for Platinum POC1-100B:
Table of Contents

Advertisement

Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
0x1120 highPressureAlarm
Internal
Fault
0x1121 lowPressureAlarm
Internal
Fault
Internal
0x1122 compressorRunningAlarm
Fault
Internal
0x1123 compressorNotRun-
ningAlarm
Fault
36
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Shut Down Internal
Fault"
The display shows:
"Shut Down Internal
Fault"
The display shows:
"Shut Down Internal
Fault"
The display shows:
"Shut Down Internal
Fault"
Possible Reason
Solution
1. Blockage in outlet
1. Remove any kinks
gas stream.
Blockage may
be internal or
2. Replace the
external.
3. If alarm persists,
1. Internal leakage of
1. Clean air intake
pneumatic circuit.
2. Blockage of inlet
2. If alarm persists,
flow.
1. Compressor or
1. If alarm persists,
compressor's
wire harness is
damaged. -or-
2. The compressor
wire harness is
damaged or loose.
1. Compressor or
1. If alarm persists,
compressor's
wire harness is
damaged. -or-
2. The compressor
wire harness is
damaged or loose.
or blockage in
cannula.
patient outlet filter.
contact Invacare
for service.
filter screen.
contact Invacare
for service.
contact Invacare
for service.
contact Invacare
for service.