General
Alarm
Alarm Code Name
Alarm
Code
Name
Battery
0x6206 LPA_batteryCold
Cold
(Running)
Unit Hot
0x5203 LPA_unitHot
(Running)
0x5204 LPA_unitCold
Unit Cold
(Running)
1187375-B
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Battery Cold Allow to
Warm"
The display shows:
"Unit Hot Allow to
Cool"
The display shows:
"Unit Cold Allow to
Warm"
Possible Reason
Solution
1. The concentrator is
1. Remove the cold
running off battery
power and at least
one of the battery
2. If alarm persists,
temperatures is
less than or equal
to –15° C
1. The concentrator is
1. Allow the unit to
running off battery
power or external
2. Clean air intake
power (AC/DC) and
the unit internal
3. If alarm persists,
temperature is
greater than or
equal to 50° C.
1. The concentrator is
1. Allow unit to
running off battery
power or external
2. If alarm persists,
power (AC/DC) and
the unit internal
temperature is less
than or equal to
5°C.
Troubleshooting
battery and allow
to warm.
contact Invacare
for service.
cool.
filter.
contact Invacare
for service.
warm.
contact Invacare
for service.
27