Invacare Platinum POC1-100B Service Manual page 33

Oxygen concentrator
Hide thumbs Also See for Platinum POC1-100B:
Table of Contents

Advertisement

General
Alarm
Alarm Code Name
Alarm
Code
Name
0x1107 SDA_stuckButton
Stuck
Button
0x110c
No Breath
SDA_noBreathDetected
Detected
(shut down)
Internal
0x1109 ricEventQueueFullAlarm
Fault
1187375-B
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Shut Down Stuck
Button"
The display shows:
"Shut Down No Breath
Detected"
The display shows:
"Shut Down Internal
Fault"
Possible Reason
Solution
1. One of the
1. Make sure no
buttons on the
concentrator's
control panel was
pressed when a
2. Disconnect all
power source is
being connected
to the unit.
3. If alarm persists,
1. An inhalation has
1. Verfy cannula is
not been detected
by the unit for
more than 120
seconds.
2. Check the outlet
3. If alarm persists,
1. PCB or software
1. If alarm persists,
fault.
Troubleshooting
control panel
buttons are
pressed.
power sources
then reconnect to
a power source.
contact Invacare
for service.
connected, not
kinked, properly
positioned and
user is breathing
through his/her
nose.
filter/Replace filter.
contact Invacare
for service.
contact Invacare
for service.
33