Invacare Platinum POC1-100B Service Manual page 31

Oxygen concentrator
Hide thumbs Also See for Platinum POC1-100B:
Table of Contents

Advertisement

General
Alarm
Alarm Code Name
Alarm
Code
Name
Battery Too
0x1102 SDA_batteryTooCold
Cold (shut
down)
Unit Too
0x1103 SDA_unitTooHot
Hot (shut
down)
Unit Too
0x1104 SDA_unitTooCold
Cold (shut
down)
1187375-B
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Shut Down Battery
Too Cold"
The display shows:
"Shut Down Unit Too
Hot"
The display shows:
"Shut Down Unit Too
Cold"
Possible Reason
Solution
1. The concentrator is
1. Remove the cold
running off battery
power and at least
one of the battery
2. Change the
temperatures is
less than or equal
3. If alarm persists,
to
–20° C.
1. The concentrator is
1. Allow unit to cool.
running off battery
2. Clean intake filter
power or external
power and the
3. If alarm persists,
unit temperature
is greater than or
equal to 55°C.
1. The concentrator is
1. Allow unit to
running off battery
power or external
2. If alarm persists,
power and the unit
temperature is less
than or equal to 0°
C.
Troubleshooting
battery and allow
to warm.
battery.
contact Invacare
for service.
screen.
contact Invacare
for service.
warm.
contact Invacare
for service.
31