Invacare Platinum POC1-100B Service Manual page 30

Oxygen concentrator
Hide thumbs Also See for Platinum POC1-100B:
Table of Contents

Advertisement

Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
0x1100 SDA_depletedBattery
Depleted
Battery
Battery Too
0x1101 SDA_batteryTooHot
Hot (shut
down)
30
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Shut Down Battery
Depleted"
The display shows:
"Shut Down Battery
Too Hot"
Possible Reason
Solution
1. The concentrator is
1. Supply unit with
running off battery
power and the
battery charge
level is less than
or equal to 5%.
2. If alarm persists:
3. If alarm persists,
1. The concentrator is
1. Remove the hot
running off battery
power and at least
one of the battery
2. Change the
temperatures is
greater than or
3. If alarm persists,
equal to 60°C.
external power
(AC/DC) or install
a charged battery
into the unit.
a. Inspect
the battery
connector for
debris.
b. Reinstall
battery.
contact Invacare
for service.
battery and allow
to cool.
battery.
contact Invacare
for service.
1187375-B

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents