Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
Low Oxygen
0x1105 SDA_lowOxygenPurity
Purity (shut
down)
32
All manuals and user guides at all-guides.com
Display Message
The display shows:
"Shut Down Oxygen
Fault"
Possible Reason
Solution
1. The concentrator
1. Ensure cannula is
is producing less
than 73% O2
beginning
2. Clean air intake
15 minutes after
the power-on
3. Change outlet
state.
a. Unit running
4. If alarm persists,
without cannula
connected -or-
b. Inlet airflow
blockage -or-
c. Outlet filter
perforated -or-
d. Sieve bed
has been
compromised
-or-
e. Oxygen sensor
is not properly
calibrated.
connected while
unit is operating.
filter screen.
filter.
contact Invacare
for service.
1187375-B