Tecfluid CU Series Instruction Manual page 30

Ultrasonic flowmeter
Table of Contents

Advertisement

When the read flow rate is stable, press the key OK and the instrument will save the value.
If in the installation is not possible to stop the flow rate, an offset adjustment can be done if
the flow rate is stable.
To do this, press the option "Flow rate ≠ 0 and the following screen will be shown:
When the read flow rate is stable, press the key OK and the instrument will save the value.
The following step is to swap the cables of the transducers: the cable TR1 for the transducer
TR1 has to be connected to the transducer TR2 and vice versa.
Do not move the transducers from its position. Only unscrew the two connectors and swap
the cables as described.
Once a key is pressed, the following screen is shown:
After pressing OK, the cables must be placed in its original position (cable TR1 to transducer
TR1 and cable TR2 to transducer TR2).