5.16

Transducers test

Through this screen a test of the transducers can be performed. This test allows to verify
that they are in good condition.
To do the test, put some gel on the transducers and face each other.
The display should indicate a value between 90% and 100%. If not, please contact the
manufacturer.
6

CONVERTER PROGRAMMING

By programming the converter the visualization and the outputs of the instrument can be
configured.
Turn on the converter and press (OK) to enter the main menu. The following screen appears.
With the keys (2) and (8), select Programming, and then validate with the key (OK).
To access the programming of the converter, you must enter the password. At first access,
the default password is 0123. For more details about the password, see paragraph 4.1,
page 12.