Siemens MOBILETT Plus Service Instructions Manual page 85

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Mechanical systems
Recess
Fig. 12
WARNING
Loosen the adjustment screw so that the spring pack is loosened and the nearest part of
the pack is lowered (Fig. 12).
WARNING
Loosen the two large Allen screws (2) and (3) on the end of the spring pack. The whole
spring pack will then fall down on the floor.
Remove the spring pack.
Install the new spring pack with the two Allen screws (2) and (3) on the end of the spring
pack.
Lift up the spring pack in the front end and check through the hole for the adjustment
screw that the front of the spring pack with recess is visible.
Mount the adjustment screw and tighten until the spring pack is held in position.
Siemens-Elema AB
Solna, Sweden
Adjustment screw
Risk of injury to the hands!
The strong tension in the spring pack could cause undesirable
actions.
Be careful during the following procedures.
If the spring pack does not lower, there is still spring tension.
The strong tension in the spring pack could cause undesirable
actions.
Check whether the single tank is in its uppermost position; if not,
raise the single tank to this position. Check whether the adjust-
ment screw is loose enough. Do not use any force to remove the
spring pack.
SPR8-215.898.02
Rev. 01
2
Page 9 of 28
01.02
SPS-UD
10 - 9
3
MOBILETT Plus/ Plus E

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus e320010900

Table of Contents