Siemens MOBILETT Plus Service Instructions Manual page 53

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

HT and single tank error diagnosis
Exposure values, 133 kV 5 mAs, HTPOS & HTNEG 2 V/div,
20 kV/div; 200 µs/div (only the end of the exposure is shown)
HTPOS on
D1
HTNEG on
D1
Fig. 6
HT-signal, Arcing between plus and minus side (Defective tube). Replace single tank.
Siemens-Elema AB
Solna, Sweden
SPR8-215.898.02
Rev. 01
01.02
Page 9 of 12
MOBILETT Plus/ Plus E
SPS-UD
6 - 9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus e320010900

Table of Contents