Siemens MOBILETT Plus Service Instructions Manual page 10

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

1 - 6
Measure the capacitor bank voltage on the bus bar (Fig. 3).
If one of the capacitor series of the capacitor bank is still charged (U > 10 V), it can be
discharged via resistor R103 on D7.
Connect the voltmeter to the measurement points parallel to the insulated cables which
will be used to discharge the capacitor bank (U < 10 V) (Fig. 4).
MOBILETT Plus/ Plus E
SPR8-215.898.02
Rev. 01
01.02
Page 6 of 6
Siemens-Elema AB
SPS-UD
General
Solna, Sweden

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus e320010900

Table of Contents