Display messages

List of displayed messages

Err 3
Err 4
Err 5
Err 6
CAL 7
___
Err 9
USE 10
___
CAL 12
CAL 13
CAL 14
USE 15
Err 16
Err 17
Err 18
CAL 19
USE 20
Err 21
Err 22
Err 23
USE 24
USE 25
CAL 26
CAL 27
CAL 28
CAL 29
USE 30
CAL 31
Err 32
___
Err 34
Err 35
Siemens-Elema AB
Solna, Sweden
Overvoltage >142 kV
Poor kV regulation
No tube current
Anode speed too slow
Exposure time too long
___
High-tension asymmetry
Single tank >50° or <8°C, measured at the location of the temperature sensor
___
Values in CMOS memory lost
Filament voltage, 60 V DC on D4 not present
Overvoltage in the capacitor bank (>360 V)
Preparation level held by the user >20 s
Capacitor bank charging too fast
Capacitor bank charging too slow
Capacitor voltage charging too slowly
Leakage current too high in the capacitor
Radiation release button was released during the exposure
Filament current incorrect for stand-by
Filament current incorrect for exposure
During exposure, the main inverter was interrupted >3 times
The selected kV and mAs values are lost
The calculated anode load is lost
Error in RAM
Error in ROM
Error in the operating system
No single tank connected or an error in the connection
The customer values stored in CMOS memory are lost
Replace backup battery <2.8 V.
Capacitance of the capacitor bank is too low
___
Error log overfull
Watchdog responded
SPR8-215.898.02
Rev. 01
01.02