Replacement Of Mains Cable/Cable Winder - Siemens MOBILETT Plus Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Miscellaneous
Fig. 2

Replacement of mains cable/cable winder

Remove side covers and the cover with the cassette compartment.
Remove the four (two on each side) screws that fixes the cable winder assembly
(pos. 1 / Fig. 2).
CAUTION
Pull out the cable winder assembly approximately 10 cm.
Disconnect main switch plug K52 (pos. 1 / Fig. 3).
Siemens-Elema AB
Solna, Sweden
1
MOB00618
Sharp edge!
The brake handle could be damaged.
Lock the brake handle, otherwise it is damaged by the sharp
upper edge of the cable winder assembly.
Fig. 3
SPR8-215.898.02
Rev. 01
01.02
Page 3 of 6
MOBILETT Plus/ Plus E
SPS-UD
8 - 3
1
8

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus e320010900

Table of Contents