Siemens MOBILETT Plus Service Instructions Manual page 29

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Display messages
Dis-
played
messages
Err 37
The mes-
sage is not
shown to
the opera-
tor. This
message
can only
be read
out from
the mes-
sage his-
tory, P08.
Err 38
The mes-
sage is not
shown to
the opera-
tor. This
message
can only
be read
out from
the mes-
sage his-
tory, P08.
PC NodE
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1
Siemens-Elema AB
Solna, Sweden
Mode of
Cause
operation
Generator ON
The error message
history is lost. It is
indicated with 0´s
and "Err 37".
Generator ON
The error statistics
are lost.
The switch S:1 on
board D1 is set in the
wrong position.
SPR8-215.898.02
Rev. 01
Possible reasons and advice
Check the supply voltage on D1,
see page 110**.
Check that the voltage in the
backup battery is > 2.8 V and that
the battery jumper is connected,
refer to page 110**.
If the message appeared once, a
tube arc may have disturbed the
program.
If the "Err 37" occurs repeatedly,
the single tank must be checked
according to section "Error
diagnosis HT and single tank". If
the function of the single tank is OK
and "USE 37" appears repeatedly,
replace D1.
See "Err 37". If "Err 38" is the only
message in the history memory, it
will not be necessary to replace
D1. These statistics are only used in
the factory.
This mode is only used at the factory.
To quit this mode; put switch S1:2 in
the OFF position.
Page 19 of 22
01.02
SPS-UD
2 - 19
MOBILETT Plus/ Plus E