Siemens MOBILETT Plus Service Instructions Manual page 22

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 12
Dis-
played
messages
Err 23
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus/ Plus E
Mode of
Cause
operation
During exposure The HT inverter has
been interrupted
several times, e.g.
overcurrents.
SPR8-215.898.02
Rev. 01
Display messages
Possible reasons and advice
When the HT inverter has been inter-
rupted more than 3 times, the expo-
sure will be interrupted.
Faulty inverter/inverter control
Check the inverter current, refer to
the section "Error diagnosis HT
and single tank".
The X-ray tube may have arced
during the exposure. If the first
exposure after a long interval is an
exposure with high kV values, tube
arcs can occur.
The unit has to be started with
several exposures at high mAs
values. The kV values selected
should be from 40 kV increased to
133 kV.
Refer also to the section "Error
diagnosis HT and single tank".
Page 12 of 22
01.02
SPS-UD
Siemens-Elema AB
Solna, Sweden