Siemens MOBILETT Plus Service Instructions Manual page 71

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Replacement and repairs on the collimator
Set a SID of 100 cm (39"). Use a measuring tape.
Switch on the collimator lamp and align the cassette. Collimate the light field to
25 x 25 cm (10" x 10").
Mark the four corners with coins and mark one of the corners with two coins (Fig. 5).
Set 52 kV, 2.0 mAs on the control panel.
CAUTION
Release an exposure.
Develop the film and measure the deviation between the light and the radiation fields.
Acceptable deviation ± 1.7% of SID, (Fig. 6).
If the deviation is more than 1.7%, adjust with the screws marked C and D (Fig. 2), to
move the light field closer to the radiation field.
Repeat the test with a new exposure.
If necessary, adjust again until the deviation between the light field and the radiation field
is within the tolerance specified.
Reattach cover 1 (Fig. 1).
Siemens-Elema AB
Solna, Sweden
X-ray radiation!
If not observed, risk for exposure to radiation.
Use radiation protection!
SPR8-215.898.02
Rev. 01
Page 3 of 8
01.02
SPS-UD
9 - 3
MOBILETT Plus/ Plus E