Siemens MOBILETT Plus Service Instructions Manual page 50

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

6 - 6
Exposure values: 90 kV 5 mAs, HT 20 kV/div (500 mV/div)
I
inverter-single tank
.
HT on
D1
Inverter
current
Fig. 3
HT and inverter current, inverter and single tank normal
MOBILETT Plus/ Plus E
HT and single tank error diagnosis
50 A/div (only the end of the exposure is shown
SPR8-215.898.02
Rev. 01
Page 6 of 12
01.02
SPS-UD
Siemens-Elema AB
Solna, Sweden