Replacement and repairs on the collimator
cable tie left side
Fig. 8
•
Disconnect the cable to the connection terminal board "D" (Fig. 8).
•
If there are cable ties, cut them (Fig. 8).
•
Remove the collimator by loosening the four screws "E" (Fig. 8). Reserve the hardware.
•
Install the new collimator. Center the rotary-flange so that the screws marked "E" are
centered in the grooves before tightening.
•
Connect the wires to connection terminal board "D".
Siemens-Elema AB
Solna, Sweden
D
E
E
SPR8-215.898.02
Rev. 01
Page 5 of 8
01.02
SPS-UD
9 - 5
E
cable tie
right side
E
MOBILETT Plus/ Plus E