Siemens MOBILETT Plus Service Instructions Manual page 16

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 6
Dis-
played
messages
Use 10
CAL 12
CAL 13
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus/ Plus E
Mode of
Cause
operation
Generator ON.
The single tank tem-
The message
perature measured
does not inter-
is > 50°C or
rupt a running
< 8°C.
exposure.
During
Stored filament
preparation
value is lost.
Generator ON
The voltage supply
(60 V DC) to the fila-
ment circuit is
beyond its limit.
SPR8-215.898.02
Rev. 01
Display messages
Possible reasons and advice
The generator is overheated and
must cool down or the generator
was stored in a cold room and must
warm up prior to use.
Too high or too low environmental
temperature.
As soon as the single tank
temperature is
< 47°C, check the measurement
circuit for the temperature; refer to
page 140**, contact K9.
Check the voltage in the backup
battery, refer to page 110**. The
battery should have a voltage
> 2.8 V
Check that the battery jumper is
securely attached, see page 110**.
Make a new adaptation, see
section "P10 - Fast adaptation".
Check fuse U6 on D8, refer to page
160**.
Check the filament supply voltage
+60 V DC, refer to page 160**.
Page 6 of 22
01.02
SPS-UD
Siemens-Elema AB
Solna, Sweden

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus e320010900

Table of Contents