Procedure - Keysight Technologies N9010A EXA Functional Tests

Hide thumbs Also See for N9010A EXA:
Table of Contents

Advertisement

Frequency Response (Flatness)

Procedure

Procedure
1. Zero and calibrate the power meter and power sensor as described in the
2. Configure the equipment as shown in
Connect the power splitter to the signal analyzer input using the appropriate
NOTE
adapter. Do not use a cable.
3. Assure the signal analyzer's built-in auto alignment has been performed
4. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer, and
5. Set up the synthesized sweeper by pressing:
6. Set up the signal analyzer by pressing:
7. Adjust the synthesized sweeper output power for a power meter reading
8. On the signal analyzer, press Single.
9. Press the Peak Search key on the signal analyzer to position the marker on
10.Refer to
11.Enter the power meter reading into the Power
12.Tune the synthesized sweeper and signal analyzer to the next frequency
13.Enter the power sensor calibration factor into the power meter.
14.For frequencies 3.6 GHz and above, press AMPTD Y Scale, then Presel
15.Repeat
Chapter 5
power meter operation manual.
within the last 24 hours.
press Preset on the synthesized sweeper.
CW, 50, MHz
Power level, –4, dBm
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, Span, 50, kHz
AMPTD Y Scale, Ref Level, 0, dBm
of –10 dBm ±0.1 dB.
the peak of the signal.
Table 5-1, "Frequency Response (Flatness) Results."
amplitude of the signal displayed on the signal analyzer into the Meas
column of
Table
5-1.
listed in
Table
5-1.
Center to center the preselector filter for an optimum amplitude
measurement.
step 7
through
for the frequency range of your analyzer.
Figure
step 14
and complete the remainder of
5-1.
Enter the
column of
Table
Meter
Table 5-1
Amp
5-1.
31