Amplitude Accuracy at 50 MHz
Procedure
Table 4-1
Amplitude Accuracy Results
Nominal
Source
Amplitude
(dBm)
−4
−9
−14
−4
−14
−4
−14
26
h. Record the signal amplitude, as measured by the analyzer in the
Measured Amplitude column of
i. Calculate the signal amplitude accuracy error using the following
equation, and record the results under the Amplitude Accuracy
Error column:
Amplitude Accuracy Error = Meas_amp − Power_meter
Attenuation
Span
(dB)
(MHz)
10
2
10
1
10
0.5
20
0.1
20
0.1
30
0.1
30
0.1
Table 4-1
Measured
Power Meter
Amplitude
Amplitude
Meas_amp
Power_meter
(dBm)
(dBm)
Amplitude
Test Limit
Accuracy
(dB)
Error
(dB)
±1.20 dB
±1.20 dB
±1.20 dB
±1.20 dB
±1.20 dB
±1.20 dB
±1.20 dB
Chapter 4