Operation Verification
Checking the Output Power

N5171B/81B Alternative Test Procedure

If a USB power sensor is not available, use a power meter to measure the
output power of the signal generator.
Test Setup
1. Zero and calibrate the power sensor to the power meter:
2. Connect the equipment as shown:
3. Preset the signal generator: Press Preset.
4. Turn RF on: Press RF On/Off so that the RF On/Off LED lights.
5. Turn modulation off: Press Mod On/Off so that the Mod On/Off LED turns
6. Set the frequency: Press Frequency and enter the first frequency value
7. Set the amplitude: Press Amplitude and enter the amplitude value for that
8. Configure the power meter for the measurement.
9. Measure the output power level.
10.Repeat steps 6 through 9 to measure power at each of the 15 frequencies
11.Confirm that the measured power levels are within the limits listed in the
Limit values are due to power meter uncertainty.
34
off.
listed in
Table
3-2.
frequency.
a. Press the Frequency Cal Fac button on the power meter.
b. Select a power meter channel (if applicable).
c. Use the arrow keys to enter the frequency at which to measure the
power.
listed in
Table
3-2.
table.
Keysight CXG, EXG, and MXG X-Series Signal Generators Getting Started Guide