Select the option: Check & adjust the compensator, index error
& line of sight error
Next
Face I measurement
Use a clean Leica standard prism as the target. Do not use
a 360° prism.
Aim the telescope accurately at a
target at about 100 m distance.
The target must be positioned
within ±9°/±10 gon of the hori-
zontal plane.
The procedure can be started in
any face.
Measure to measure and to con-
tinue to the next screen.
Non-motorised instruments guide
to the other face.
The fine pointing has
to be performed man-
ually in both faces.
Face II measurement
Measure to measure the same target in the other face and to calcu-
late the instrument errors.
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and none of them is averaged with the results from previous
runs.
Adjustment Status
Number of measurements: Shows the number of runs completed.
One run consists of a measurement in face I and face II.
l Component quality (1 σ): and similar lines show the standard
deviations of the determined adjustment errors. The standard devia-
tions can be calculated from the second run onwards.
Measure at least two runs.
Next to continue with the check & adjust procedure.
± 9°
0016433_001
180°
180°
0016434_001
Check & Adjust