KYORITSU 4116A Instruction Manual And Users Manual

KYORITSU 4116A Instruction Manual And Users Manual

Digital psc-loop tester

Advertisement

Quick Links

INSTRUCTION  MANUAL
DIGITAL PSC-LOOP TESTER
MODEL 4116A, 4118A, 4120A
KYORITSU  ELECTRICAL  INSTRUMENTS
WORKS, LTD.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4116A and is the answer not in the manual?

Questions and answers

Summary of Contents for KYORITSU 4116A

  • Page 1 INSTRUCTION  MANUAL DIGITAL PSC-LOOP TESTER MODEL 4116A, 4118A, 4120A KYORITSU  ELECTRICAL  INSTRUMENTS WORKS, LTD.
  • Page 2: Table Of Contents

    CONTENTS 1. SAFE TESTING ................  1 2. PROCEDURE OF REMOVING COVER .........  3 3. FEATURES ..................  4 3.1  Instrument Layout ..............  4 3.2  Test Lead ..................  5 3.3  Features ...................  5 4. SPECIFICATIONS ................  7 5. OPERATING INSTRUCTIONS ............  8 5.1 Initial Checks ................  8 5.2 Measurement of the Loop Impedance ........  9 5.3 Measurement of Prospective Short Circuit Current ....  10 6. DETAILED EXPLANATION ............  10 6.1 Measurement of Fault Loop Impedance and Prospective Fault Current ............  10 6.2 Measurement of OLD-TT System ..........  15 6.3 Measurement of Line Impedance and  Prospective Short Circuit Current ..........  16 7. SERVICING ...................  17 8. CASE AND STRAP BELT ASSEMBLY .........  17...
  • Page 3: Safe Testing

    Therefore,  read  through  these  operating  instructions  before using the instrument. IMPORTANT: 1. This  instrument  must  only  be  used  by  a  competent  and  trained person  and  operated  in  strict  accordance  with  the  instructions. KYORITSU  will  not  accept  liability  for  any  damage  or  injury  caused by misuse or non-compliance with the instructions or with the safety procedures. 2. It is essential to read and to understand the safety rules contained in the instructions or with the safety procedures.
  • Page 4 ● Never attempt to use the instrument if the instrument or your hand is wet. CAUTION ● For  those  testers  without  the  D-LOK  circuitry  (models  4116A, 4118A) all RCD's (RCCB, ELCB) in the circuit must be by-passed for the duration of the test    (except on loop-2000Ω range). Do not operate the RCD Test Button with the RCD by-passed. ● During  testing  it  is  possible  that  there  may  be  a  momentary degradation  of  the  reading  due  to  the  presence  of  excessive transients ...
  • Page 5: Procedure Of Removing Cover

    2. PROCEDURE OF REMOVING COVER Model 4116A, 4118A and 4120A have a dedicated cover to protect against an impact from the outside and prevent the operation part, the LCD and the connector socket from becoming dirty. The cover can be detached and put on the back side of the main body during measurement. 2.1 Method of removing the cover Fig. 1 2.2 Method of housing the cover Fig. 2 ― 3 ―...
  • Page 6: Features

    3. FEATURES 3.1 Instrument Layout ⑤ ① ② Wiring correct (Green LED's) Reversed L/N terminals (Red LED) ③ ④ Connector Fig. 3 1.... LCD display 2.... Wiring check LEDs  3...  Test button  4.... Range switch 5.... Connector DANGER ● Use original test lead only. ● Max. allowed voltage between mains test terminals and ground is 300V. ● This instrument is intended only for use in single phase operation at  230V  +10%  -15%  AC  phase  to  earth  or  for  use  in  OLD-TT system phase to neutral. ―...
  • Page 7: Test Lead

    3.2 Test Lead The instrument is supplied with Model 7125 lead at socket outlets and Model 7121 distribution board lead. (1) Model 7125 (2) Model 7121 (Included as standard equipment for M-4118A and 4120A. Optional for M-4116A.) Black-Neutral Red-Phase Green-Earth Fig.4 3.3 Features 3.3-1 Test Range (Function): Model M-4120A M-4118A M-4116A ○ × × D-LOK Circuit * Loop 0-19.99Ω/ ○ ○ ○ 0-199.9Ω/0-1999Ω PSC 0-199.9A/ ○ ○ × 0-4.00kA 0-1999A/ Note: ● *D-LOK=Automatic device (RCD) lock ● D-LOK dose not operate on the Loop-2000Ω. ● Mains voltage with which D-LOK operates is as shown in the table below. Range D-LOK Operating Voltage Loop200Ω/PSC200A...
  • Page 8 3.3-3 Model 4116A, 4118A and 4120A have the following features: ● Battery is not used All models are not battery-operated, but operate by the voltage supplied from the system. ● Wiring check Three  LEDs  indicate  if  the  wiring  of  the  circuit under  test  is  correct.  The  P-E  and  P-N  LEDs illuminate  when  the  wiring  polarity  of  the  circuit under test is correct. The        Reverse  LED  is  lit  when  P  and  N  are reversed.
  • Page 9: Specifications

    20kA 0.00 〜 4.00kA 25A / 20ms Voltage Measuring range Accuracy 110 〜 260V ±(2%rdg + 4dgt) Instrument dimensions 186×167×89mm Instrument weight M-4120A / 960g, M-4116A and M-4118A / 750g Reference conditions Specifications are based on the following conditions except where otherwise stated: 1. Ambient temperature:  23 ± 5 ℃ 2. Relative humidity:45% to 75% 3. Position: horizontal 4. AC power source: 230V, 50Hz 5. Altitude: Up to 2000m Operating temperature  0  to  +40℃,  relative  humidity  80%  or  less,  no and humidity condensation.
  • Page 10: Operating Instructions

    ● Operating Error of Loop Impedance (61557-3) Measuring range to keep Maximum percentage  Range operating error operating error 20Ω 0.20 〜 19.99Ω 200Ω ±30% 20.0 〜 199.9Ω 2000Ω 200 〜 1999Ω The  influencing  variations  used  for  calculating  the  operating  error  are denoted as follows:   Temperature:0℃ and 40℃ Phase angle :At a phase angle 0゜to 18゜ System frequency:49.5Hz to 50.5Hz System voltage:230V+10%-15% 5. OPERATING INSTRUCTIONS 5.1 Initial Checks  - To be carried out before any testing. (1) Test Lead Connection Insert  the  lead  plug  into  the  connector  on  the  instrument  correctly  as shown below.
  • Page 11: Measurement Of The Loop Impedance

    WARNING ● If the above sequence is NOT displayed or the RED LED is on for any  reason  ,  DO  NOT  PROCEED  AS  THERE  IS  INCORRECT WIRING. The cause of the fault must be investigated and rectified. (3) Voltage Measurement When the instrument is first connected to the system, it will display the phase-neutral voltage which is updated every 1s. This mode is cancelled whenever  the  test  button  is  pressed.  If  this  voltage  is  not  normal  or  as expected, DO NOT PROCEED. WARNING ● This instrument is intended only for use in single phase operation at  230V  +10%  -15%  AC  phase  to  earth  or  for  use  in  OLD-TT system phase to neutral.
  • Page 12: Measurement Of Prospective Short Circuit Current

    5.3 Measurement of Prospective Short Circuit Current  − (Models 4118A and 4120A) (1) Set the instrument to the 20kA range. (2) Connect the test lead to the instrument. (3) Attach the plug to the socket to be tested. (4) Check that the LED's are lit in the sequence indicated in section 5.1. If not, disconnect from the mains and check the wiring at the socket. (5) Press the "Press to Test" button. The prospective short circuit current (PSC) will be directly displayed on the LCD with the appropriate units. This will remain for 3s and then revert to AC voltage display. An audible beep will sound on completion of the test. For best results always test on the lowest possible range. For  example  a  PSC  measured  on  the  2000A  range  may  read  60A whilst on the 200A range it may read 56.0A. To hold the reading keep the button held down or turn clockwise to lock for Auto Test. Note: ● For loop impedance greater than 210Ω on PSC 200A range and 25 Ω on 2000A, 20kA ranges, the fault voltage may become high and dangerous due to the D-LOK current, therefore, the unit is designed to lock out PSC ranges showing "Uf-Hi" symbol. Normally PSC tests are conducted at point of origin, e.g. distribution boards, between phase and neutral. When  conducting  PSC  tests  at  socket  outlets,  a  test  will  be  conducted between phase and earth due to the fixed wiring of the moulded mains plug.
  • Page 13 In  the  event  of  a  fault,  the  Fault  loop  impedance  should  be  low  enough (and  the  Prospective  Fault  current  higher  enough)  in  order  to  have  the automatic disconnection of supply by the installed protection device within prescribed time interval. Every circuit must be tested to make sure that the fault loop impedance does  not  exceed  that  specified  for  the  over  current  protection  device concerned.
  • Page 14 ⊿ For instance in a TT system protected by a RCD the max RA values are: Rated residual operating current I⊿n. 1000 RA (at 50V) 5000 1667 Ω 2500 Ω RA (at 25V) Note: ● The  loop  tester  models  4120A  /  4118A  /  4116A  measure  the  fault loop impedance that is a value normally a little bit higher of RA. But,  if  the  electrical  installation  is  protected  considering  the  loop impedance value, also the RA formula will be fulfilled. ― 12 ―...
  • Page 15 Practical example of verification of the protection in a TT system according to the international Standard IEC 60364. RCD 30mA Fig.8 For  this  example  max  value  is  1667Ω,  the  loop  tester  reads  12.74Ω,  it means that the condition RA < 50/Ia is respected. It  is  fundamental  for  this  example  to  test  also  the  RCD  to  ensure  that operation takes place quickly enough to respect the safety requirements. In order to do it, can be used the RCD tester model 5406A. According  to  the  international  Standard  IEC  60364  for  TN  system  the following condition shall be fulfilled for each circuit: <...
  • Page 16 For  instance  in  a  TN  system  with  nominal  mains  voltage  Uo  =  230  V protected by gG fuses the Ia and max Zs values could be: Disconnecting time 5s Disconnecting time 0.4s Rating  Ia  (A) Zs  (Ω) Ia  (A) Zs  (Ω) 1.56 1.25 1.53 0.83 1.21 0.72 0.92 0.49 0.71 0.42 0.54 0.27 0.39 1020 0.22 Using the current ranges on models 4120A and 4118A can be also tested the Prospective Fault current. ...
  • Page 17: Measurement Of Old-Tt System

    Max value of Zs for this example is 2.1Ω (16A gG fuse, 0.4s) the loop tester reads 1.14Ω (or 202 A on Fault current range) it means that the condition Zs < Uo/Ia is respected. In fact the Zs of 1.14Ω is less than 2.1Ω (or the Fault current of 202 A is more than Ia of 110A). WARNING ● This instrument is intended only for use in single phase operation at  230V  +10%  -15%  AC  phase  to  earth  or  for  use  in  OLD-TT system phase to neutral. ● If the overheat symbol appears in the display (  ) disconnect the instrument from the mains supply and allow to cool down. 6.2 Measurement of  "OLD-TT System" OLD-TT  system  is  a  TT  system  with  phase  to  phase  voltage  of  220  V (instead ...
  • Page 18: Measurement Of Line Impedance And Prospective Short Circuit Current

    WARNING (OLD-TT system only) ● DO NOT PRESS the "Test button" if the display reads a value of 220V! The RCD could trip during the tests also with model 4120A.  CAUTION ● The  D-LOK  circuit  of  4120A  can  not  works  with  127V  between phase to earth. 6.3 Measurement of Line Impedance and Prospective Short Circuit Current Line  Impedance  on  single-phase  system  is  the  impedance  measured between phase and neutral terminals. Measurement principle used inside the instrument is exactly the same as at Fault  Loop  Impedance  measurement,  but  the  measurement  is  carried  out between L and N terminals.
  • Page 19: Servicing

    WARNING ● This instrument is intended only for use in single phase operation at  230V  +10%  -15%  AC  phase  to  earth  or  for  use  in  OLD-TT system phase to neutral. ● If the overheat symbol appears in the display (  ) disconnect the instrument from the mains supply and allow to cool down. ● When testing installation that has a large current capacity, such as a power line, be sure not to short live conductors with the tip of a probe. Failure to follow these instruction can cause hazards to the user. 7. SERVICING If this tester should fail to operate correctly, return it to your distributor stating the exact nature of the fault.  Please remember to give all the information possible concerning the nature of  the  fault,  as  this  will  mean  that  the  instrument  will  be  serviced  and returned to you more quickly.
  • Page 20 DISTRIBUTOR 92-1483 01-05 Kyoritsu  reserves  the  rights  to  change  specifications  or  designs described in this manual without notice and without obligations.

This manual is also suitable for:

4118a4120a

Table of Contents