The Biopsy Procedure - GE Versana Essential Technical Publication

Table of Contents

Advertisement

The Biopsy Procedure

WARNING
NOTE:
Versana Essential
5759238-100 English Rev. 2
Biopsy procedures must only be performed on live images.
1. Place coupling gel on the scanning surface of the probe/
sheath/biopsy guide assembly.
2. Activate the biopsy guidezone on the system through the
B-Mode menu. When using multi-angle guides, ensure that
the proper guidezone angle is displayed.
3. Scan to locate the target. Center the target in the electronic
guidezone path.
Enabling color flow would allow for visualization of the
vascular structure around the area to be biopsied.
4. Place the needle in the guide between the needle barrel and
needle clip. Direct it into the area of interest for specimen
retrieval.