GE Versana Essential Technical Publication page 229

Table of Contents

Advertisement

Notice upon Installation of Product
Versana Essential
5759238-100 English Rev. 2
Separation distance and effect from fixed radio communications
equipment: field strengths from fixed transmitters, such as base
stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV
broadcast transmitter cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in
which the ultrasound system is used exceeds the applicable RF
compliance level as stated in the immunity declaration, the
ultrasound system should be observed to verify normal
operation. If abnormal operation is observed, additional
measures may be necessary, such as re-orienting or relocating
the ultrasound system or using an RF shielded examination
room may be necessary.
1. Use either power supply cords provided by GE or ones
designated by GE. Products equipped with a power source
plug should be plugged into the fixed power socket which
has the protective grounding conductor. Never use any
adaptor or converter to connect with a power source plug
(e.g. three-prong-to-two-prong converter).
2. Locate the equipment as far away as possible from other
electronic equipment.
3. Be sure to use only the cables provided by or designated by
GE. Connect these cables following the installation
procedures (e.g. wire power cables separately from signal
cables).
4. Lay out the main equipment and other peripherals following
the installation procedures described in the service manuals
and peripherals manufacture's manuals.
User Guide
Safety Precautions
4-21