ERROR CONDITIONS

If the Printer/Charger should detect an error, it will generate an
audible alarm and flash the DATA light in groups of
which correspond to the error that was detected:
FLASH
There is no paper in the printer, or the printer lever is
1
-
l
in the wrong (forward) position.
2 FLASHES
l
STOP button to clear the problem. If the error recurs, the
Printer/Charger may need servicing.
3 FLASHES - Data received from the PneumoCheck is invalid.
l
Press STOP, and retransmit data by removing the Pneumo-
Check from the Printer/Charger well temporarily. If the error
recurs, turn the PneumoCheck OFF, clear the Printer/Charger by
pressing STOP again, and retest the patient. If the problem
persists, your PneumoCheck, Printer/Charger, or both, may
require servicing.
NOTE: The version of software in your PneumoCheck Printer/
Charger is printed on each printout just beneath the chart of values.
ACTUAL
NORMS: Knudson 1983
G o o d FVC maneuver
A hardware malfunction has occurred. Press the
-
NORM %NORM ABN
30
2, or 3 flashes
1,
VERSION