Message
PROBE NOT CLEAR
Please ensure the probe is not
blocked or obstructed
AIRFLOW ERROR
Unknown pump fault. Restart the
unit. If problem persists, contact
Welch Allyn
WARNING! CALIBRATION EXPIRED.
Recalibration needed before further
tests are performed
"WARNING! BATTERIES LOW.
Recharge the batteries before
performing tests
Powering down
AIRFLOW ERROR.
Cannot determine pump direction. If
DIR 80024157 Ver A
C
AULT
ONDITIONS
In some fault conditions the internal pump may
Meaning / Action
Examine the probe tip for blockages. If necessary,
remove it and clean or replace it. If the problem
persists, contact your Welch Allyn service center.
The current date is later than the next calibration
date. Check that the clock is set to the correct date. If
so, arrange for the instrument to be recalibrated.
Tests can still be performed.
Recharge the batteries immediately
Other than after the specified power off delay, the
MicroTymp 4 may turn off because the internal
batteries are spent. To replace the batteries, contact
your Welch Allyn service center.
Pump fault. If the fault persists, contact your Welch
Allyn service center.