BATTERY MAINTENANCE

The PneumoCheck will remain charged and ready for use if returned
to Printer/Charger when not in use.

BATTERY RECHARGING

The nickel cadmium batteries in the PneumoCheck will provide
approximately 8 hours of continuous operation on a full charge.
When the batteries are low, the message LOW BATT will be dis-
played, and the PneumoCheck will not allow any tests until the
batteries have been recharged.
NOTE: After LOW BATT appears, printing of the most recent test in
the PneumoCheck memory is still possible.
A fully drained battery may be fully recharged in about 15 hours by
either of these methods:
PRINTER/CHARGER
1.
Connect the power plug of the Printer/Charger to a receptacle of
appropriate voltage, frequency and plug type. The orange power
indicator will become illuminated.
2.
Place the PneumoCheck into the charging well with the display
facing you. When properly seated, the orange charging indicator
on the PneumoCheck will become illuminated.
37