Bed Exit Alarm

The nurse call is always active no matter what selection is made. When selecting nurse call only,
the audio alert will be disabled in the room. This unit will chime on an alarm if a voice message
has not been recorded and the unit is set to Voice Recording.
1. Access the Menu by pressing the UP or DOWN arrow until Bed
Exit Settings is highlighted and then press ENTER.
2. Select Set Alert Type and then press ENTER
3. Choose one of the four alert types and then press ENTER. The selection has now been
made and the user will be returned to the Home Screen.

Scale System: Zeroing the Scale

With the patient out of the bed, press and hold the ZERO button for 3 seconds. This process sets
the tare weight (base weight) and will return to the Home Screen when complete.
With the tare weight set, the patient may now be placed in the bed and begin recording weight.
NOTE: Make sure all equipment and linens are on the bed before zeroing the scale. If an air
support mattress is being used with the patient, be sure to fully inflate the mattress prior to
zeroing the scale. Adding or removing items during the zeroing process will cause inaccurate
patient weight. It is highly recommended that the scale be zeroed before each new patient.