Hantek 6052BE User Manual

Hantek 6052BE User Manual

Pc usb digital oscilloscope

Advertisement

Quick Links

PC USB DIGITAL
OSCILLOSCOPE
USER'S MANUAL
Hantek 6052BE/6082BE
www.hantek.com

Advertisement

Table of Contents
loading

Summary of Contents for Hantek 6052BE

  • Page 1 PC USB DIGITAL OSCILLOSCOPE USER’S MANUAL Hantek 6052BE/6082BE www.hantek.com...
  • Page 2: Table Of Contents

    Content General Safety Summary ................1   Chapter 1 Getting Start .................. 3   1.1 System Requirement ......................4  1.2 Install Software ......................... 5  1.3 Install Drive ........................8  1.4 General Check ......................... 13  1.5 Probe Compensation ....................... 13  1.6 Self Calibration ....................... 15  1.7 Function Check .......................
  • Page 3: General Safety Summary

    Hantek 6052BE/6082BE General Safety Summary Review the following safety precautions carefully before operate the device to avoid any personal injuries or damages to the device and any products connected to it. To avoid potential hazards use the device as specified by this user’s guide only.
  • Page 4 Hantek 6052BE/6082BE Safety Information In order to ensure the correct using and the best efficient service, please carefully read the user’s manual. System Symbols These symbols may appear in this manual or on the instrument. Warning: “Waring” identifies conditions and actions that pose hazards to the users.
  • Page 5: Chapter 1 Getting Start

    Hantek 6052BE/6082BE Chapter 1 Getting Start The oscilloscope is small, lightweight, no external power required, portable oscilloscopes. The oscilloscopes is ideal for production test, research and design and all of the applications involving analog circuits test and troubleshooting, as well as education and training.
  • Page 6: System Requirement

    Hantek 6052BE/6082BE 1.1 System Requirement To run the oscilloscope software, the needs of computer configuration are as follows: Minimum System Requirements Operating System Windows XP/Vista/Win 7 Processor Upwards of 1.00G processor Memory 256M byte Disk Space 500M disk free space...
  • Page 7: Install Software

    Hantek 6052BE/6082BE 1.2 Install Software Caution: You must install the software before using the oscilloscope. 1. While in Windows, insert the installation CD into the CD-ROM drive. 2. The installation should start up automatically. Otherwise in Windows Explorer, switch to the CD-ROM driver and run Setup.exe.
  • Page 8 Hantek 6052BE/6082BE 4. Choose a destination directory. Click ‘Next’ to continue. 5. Check the setup information. Click Next to start copying of files. 6. This Status dialog is displayed during copying of files. USER’S MANUAL...
  • Page 9 Hantek 6052BE/6082BE 7. Updating your system configuration. 8. The installation is complete USER’S MANUAL...
  • Page 10: Install Drive

    Hantek 6052BE/6082BE 1.3 Install Drive Example: Hantek6052BE 1. Connect the A-Type Plug of USB cable to your PC’s USB port. 2. Connect the B-Type Plug of USB cable to Hantek6052BE USB port. 3. New hardware is found. USER’S MANUAL...
  • Page 11 Hantek 6052BE/6082BE 4. New hardware search wizard starts. 5. New hardware search wizard starts to search the drive. USER’S MANUAL...
  • Page 12 Hantek 6052BE/6082BE 6. New hardware wizard installs software 7. Finish new hardware search wizard. USER’S MANUAL...
  • Page 13 Hantek 6052BE/6082BE 8. New hardware is found. 9. Found new hardware Wizard. USER’S MANUAL...
  • Page 14 Hantek 6052BE/6082BE 10. New hardware search wizard starts to search. 11. New hardware wizard installs software and finish new hardware search wizard. USER’S MANUAL...
  • Page 15: General Check

    Hantek 6052BE/6082BE 12. Your new hardware is installed and ready to use. 1.4 General Check Please check the instrument as following steps after receiving an oscilloscope: Check the shipping container for damage: Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
  • Page 16 Hantek 6052BE/6082BE Correctly Compensated Over compensated Under Compensated If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the fattest square wave being displayed on the oscilloscope. Repeat if necessary. WARNNING: To avoid electric shock while using the probe, be sure the perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source.
  • Page 17: Self Calibration

    Hantek 6052BE/6082BE .6 Self Calibration The self calibration routine lets you optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at any time but you should always run the routine if the ambient temperature changes by 5v or more. For accurate calibration, power on the oscilloscope and wait twenty minutes to ensure it is warmed up.
  • Page 18 Hantek 6052BE/6082BE 2. Set the CH1 probe attenuation of the oscilloscope to X10. (The default is X1). 3. Attach the tip of probe and ground nip to the Connector of Probe compensator. Click button. A square wave will be displayed within a several seconds.
  • Page 19: Accessories

    Hantek 6052BE/6082BE 1.8 Accessories All the accessories listed below are standard accessories for the oscilloscope: Probe x 2 (1.5m), 1:1, (10:1) Passive Probes A User’s Guide An USB Cable A PC software of the oscilloscope USER’S MANUAL...
  • Page 20: Chapter 2 Operating Basics

    Hantek 6052BE/6082BE Chapter 2 Operating Basics The User’s Interface The Menu System The Horizontal System The Vertical System The Trigger System Input Connectors USER’S MANUAL...
  • Page 21: The User's Interface

    Hantek 6052BE/6082BE 2.1 The User’s Interface Click the software icon on the desk after you finished the software setting and equipment connecting. Then a user interface will be showed as follows: 1. The Main Menu All setting can be found in the main menu.
  • Page 22 Hantek 6052BE/6082BE 10. Displays the CH2 information Readout show the coupling of the channels. Readout show the vertical scale factors of the channels. A “B” icon indicates that the channel is bandwidth. 11. Displays the CH1 information Readout show the coupling of the channels.
  • Page 23: The Menu System

    Hantek 6052BE/6082BE 2.2 The Menu System The Main Menu: 1. File: Load or save CSV, waveform and setup 2. View: Change the user interface 3. Channel: Set the vertical system 4. Trigger: Change trigger Mode, Sweep, Source, Slope and so on.
  • Page 24 Hantek 6052BE/6082BE 5. Horizontal: Sets time base parameters and format parameters. 6. Cursor: Set cursor measure type 7. Measure: Set measurement parameters 8. Display: Change wave display type and color. 9. Acquire: Start, Stop, Autoset and other operation setting USER’S MANUAL...
  • Page 25 Hantek 6052BE/6082BE 10. Utility: Utility setting 11. Help: Turn on help file USER’S MANUAL...
  • Page 26: The Horizontal System

    Hantek 6052BE/6082BE 2.3 The Horizontal System Click “Horizontal” in main menu. The following figure shows the horizontal parameters setting. 1. Time/DIV: leads the setting of the time base parameters 2. Format: leads the setting of the horizontal format parameters Of course, the user can also set the parameters in the right sidebar.
  • Page 27: The Trigger System

    Hantek 6052BE/6082BE 1. Select channel: 2. VLTS/DIV: Set the selected channel voltage range. 3. Coupling: Set the selected channel to DC, AC and GND. 4. BW Limit: Reject the frequency component higher than 20MHz. 2.5 The Trigger System Click “Trigger” in main menu.
  • Page 28 Hantek 6052BE/6082BE 1. Trigger Mode: Sets the trigger mode. 2. Trigger Sweep: Selects the trigger sweep mode to AUTO, NORMAL or SINGLE. 3. Trigger Source: Selects the trigger source to CH1, CH2, ALT, EXT. (Note: Hantek6082BE: CH1, CH2, ALT, EXT., EXT./10) 4.
  • Page 29: Input Connectors

    Hantek 6052BE/6082BE 2.6 Input Connectors CH 1/CH 2: Input connectors for waveform display. EXT.: Input connector for an external trigger source. Use the Trigger Menu to select the Ext. source. Other Connector: GND.: Ground terminal USB PORT: Connect the B-Type Plug of USB cable to this port.
  • Page 30: Chapter 3 Functions

    Hantek 6052BE/6082BE Chapter 3 Functions Set Oscilloscope Set Vertical System Set Horizontal System Set Trigger System Measure Signal Utility Function The Display System Save/Load Prin USER’S MANUAL...
  • Page 31: Setup The Oscilloscope

    Hantek 6052BE/6082BE 3.1 Setup the Oscilloscope Use “AUOSET” to display a signal automatically. Auto setup functions one time each time you push the “AUTOSET” button. The function obtains a stable waveform display for you. It automatically adjusts the vertical scale, horizontal scale and trigger settings.
  • Page 32: Set Vertical System

    Hantek 6052BE/6082BE 3.2 Set Vertical System Click “Channel” in main menu. 3.2.1 Change CH1/CH2 Click “CH1/CH2”in “Channel” menu. The CH1 Menu: The CH2 Menu: USER’S MANUAL...
  • Page 33 Hantek 6052BE/6082BE The CH1/CH2 control panel in sidebar: The Vertical function: Turn ON/OFF: Turn on/off the channel Volt/DIV: Select the channel voltage/div Coupling: Select the channel coupling Probe: Select the channel probe attenuation Filter: Select software filter Reset: Set the channel vertical position to zero Invert: Turn on/off the invert function.
  • Page 34 Hantek 6052BE/6082BE You can also change “CH1/CH2” voltage in side bar. You cam left click and drag the mouse on the green knob or the yellow knob to change the voltage. Set Channel Coupling Click “Coupling” in “Channel” menu. You can set the coupling to DC, AC or GND. If you set the coupling to DC, it blocks the AC component of the input signal.
  • Page 35 Hantek 6052BE/6082BE The probe setting window in the sidebar: Caution: The attenuation factor changes the vertical scale of the oscilloscope so that the measurement results refect the actual voltage levels at the probe tip. Invert The invert function turns the displayed waveform 180 degrees, with respect to the ground level.
  • Page 36 Hantek 6052BE/6082BE The following picture shows the waveform before inversion: The following picture shows the waveform of inversion: Set the Channel Bandwidth Limit The oscilloscope is set to full bandwidth and will pass the high frequency component in the signal if the “BW Limit” was turned off.
  • Page 37 Hantek 6052BE/6082BE The oscilloscope will reject the frequency component higher than 20MHz if the “BW Limit” was turned on. When the “BW Limit” was turned on, a “B” sign will be displayed at the bottom of display screen. Filter definition: None: without the Filter, read out the actual value directly.
  • Page 38 Hantek 6052BE/6082BE Source A/B: Source A and Source B menu: Operate: Four Types: A + B Add source A and source B A - B Subtract source B from source A A × B source A Multiply source B A / B...
  • Page 39 Hantek 6052BE/6082BE The Math Function Display: 3.3.3 Set Reference Click “Reference” in “Channel” menu to set REF channel. The Reference Channel Function: Turn On/Off: Turn on/off the reference channel. Volt/DIV: Channel the resolution of the reference channel. Reset: Reset the reference channel vertical position to zero.
  • Page 40 Hantek 6052BE/6082BE The Reference Waveform Display Window: Caution: If you turn on the “Reference” channel, the load file window will appear. USER’S MANUAL...
  • Page 41: Set Horizontal System

    Hantek 6052BE/6082BE 3.3 Set Horizontal System Click “Horizontal” in main menu. 3.3.1 Change Time/Div The “Time/Div” Menu: Select the horizontal Time/Div (scale factor) for the main or the window time base. The Horizontal Panel Click the blue knob can change Time/Div.
  • Page 42 Hantek 6052BE/6082BE In the “Format” item, set the waveform display format (Y-T, X-Y and Roll). Y -T: Shows the relative relation between vertical voltage and horizontal time. Roll: In Roll Mode, the waveform display updates from right to left. X -Y: Show CH1 value at X axis; CH2 value at Y axis.
  • Page 43: Set Trigger System

    Hantek 6052BE/6082BE 3.4 Set Trigger System 3.4.1 Set Edge Trigger Click “Trigger” in main menu: The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms.
  • Page 44 Hantek 6052BE/6082BE Source: You can use the trigger source options to select the signal that the oscilloscope uses as a trigger. The source can be any signal connected to a channel BNC, or to the EXT. BNC. CH1: Select CH1 as trigger signal...
  • Page 45 Hantek 6052BE/6082BE 3.4.2 HF Rejection Click “HF Rejection” in “Trigger” menu: The user can turn on “HF Rejection” to eliminate trigger higher-frequency (20M above). USER’S MANUAL...
  • Page 46: Save/Load

    Hantek 6052BE/6082BE 3.5 Save/Load Click “File” in main menu to save waveform, setups and screen. Save: Save CSV Save waveform data as a CSV file 2. Save Image Save the software display window as a .bmp or .jpg file 3. Save Setups Save the current oscilloscope setups to file 4.
  • Page 47: Utility Function

    Hantek 6052BE/6082BE 3.6 Utility Function Click the “Menu -> Utility” to get into the “Utility” menu. The “Utility” Functions: Record Pass/Fail Factory Setup Language USER’S MANUAL...
  • Page 48 Hantek 6052BE/6082BE 3.6.1 Fast Fourier Transform Function You can use the FFT mode to convert a time-domain signal into its frequency components (spectrum). You can use the FFT mode for the following types of analysis: To use the FFT mode, you need to click “Menu->Utility->FFT” to open the FFT setup window.
  • Page 49 Hantek 6052BE/6082BE The windows reduce spectral leakage in the FFT spectrum. No. of Freq Bins: Selects the FFT number Scale: Selects the vertical scale units Vertical: Selects the vertical scale Horizontal: Selects the FFT zoom factor Sets horizontal scale, If the information is more than the quantity displayed to the display window, more information is displayed by using scroll bar.
  • Page 50 Hantek 6052BE/6082BE The FFT display Windows: USER’S MANUAL...
  • Page 51 Hantek 6052BE/6082BE 3.6.2 Record and Play Back Click “Record” in “Utility” menu. The Record window will display. The following picture shows the Record Interface. This function can record input waveform form CH1 or CH2. The maximum record length is 1000 frames.
  • Page 52 Hantek 6052BE/6082BE ource: Select record source channel. (CH1 or CH2) End Frame: Set the number of record times. The max frames are 1000. Time Interval: Set the time interval of record waveform. Record: Record counter, it shows the record frames.
  • Page 53 Hantek 6052BE/6082BE End Frame: Set the end frame of play back. Current Frame: Displays the current frame of play back. You can also change this number to watch the waveform one by one. Repeat: If it was checked, the waveform will be played back repeatedly.
  • Page 54 Hantek 6052BE/6082BE The Pass/Fail function monitors changes of signals and outputs pass or fail signals by comparing the input signal with the pre-created mask. Control Setting Source: Select the Pass/Fail channel Output: Select the Pass/Fail output condition Stop When Output: If it was checked, the Pass/Fail will stop when output.
  • Page 55 Hantek 6052BE/6082BE Mask Setting Vertical: Set the vertical limit range Horizontal: Set the horizontal limit range “Create” button: Click this button to create Pass/Fail area according to the mask “Save” button: Click this button to save the setups to file.
  • Page 56 Hantek 6052BE/6082BE lick “Start” button to start the Pass/Fail test. Click “Stop” button to stop the Pass/Fail test. The Pass/Fail function display: NOTE:Pass/Fall function is unavailable in X-Y mode and Roll mode. 3.6.4 Factory Setup Click “Factory Setup” in “Utility” menu to load default setups.
  • Page 57: Measure Signal

    Hantek 6052BE/6082BE The Factory Setup function does not reset the following settings: Language option Date and time 3.6.5 Language Click “Language” in “Utility” menu. There are four languages in “Language” menu. The default language is English. 3.7 Measure Signal 3.7.1 Cursor Click “Cursor”...
  • Page 58 Hantek 6052BE/6082BE use Trace type. When you use cursors, be sure to set the Source to the waveform on the display that you want to measure. 2. Type There are four types of cursors: Cross, Trace, Vertical and Horizontal. 1) Cross...
  • Page 59 Hantek 6052BE/6082BE The Trace cursor display window: The Trace cursor measure result display on status bar Vertical The Vertical cursors appear as vertical lines on the display and measure the vertical parameters. The Vertical cursor display window: USER’S MANUAL...
  • Page 60 Hantek 6052BE/6082BE The Vertical cursor measure result display on status bar: Horizontal The Horizontal cursors appear as horizontal lines on the display and measure the horizontal parameters. The Horizontal cursor display window: The Horizontal cursor measure result display on status bar: 3.7.2 Measure Menu...
  • Page 61 Hantek 6052BE/6082BE Source The user can use the “Source” menu to select a measure source. Vertical Maximum: Voltage of the absolute maximum level, measured over the entire waveform Minimum: Voltage of the absolute minimum level, measured over the entire waveform...
  • Page 62 Hantek 6052BE/6082BE Horizontal Period: Time to take for the signal cycle to complete in the waveform Frequency: Reciprocal of the period of the first cycle in the waveform Rise Time: Time taken from lower threshold to upper threshold Fall Time: Time taken from upper threshold to lower threshold...
  • Page 63: The Display System

    Hantek 6052BE/6082BE Clear Measure Clear all measure items on display screen. The Measure Display Window: Caution: The results of the automatic measurements will be displayed on the bottom of the screen. Maximum 8 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left, out of screen.
  • Page 64 Hantek 6052BE/6082BE If the Vectors type mode is selected, the waveform will be displayed as following figure. If the Dots type mode is selected, the waveform will be displayed as following figure. USER’S MANUAL...
  • Page 65 Hantek 6052BE/6082BE 3.8.2 Display Grid Click “Display” in main menu. The grid shows: USER’S MANUAL...
  • Page 66 Hantek 6052BE/6082BE Without the grid shows: USER’S MANUAL...
  • Page 67 Hantek 6052BE/6082BE 3.8.3 Intensity Click “Display->Intensity” In main menu The following figure shows the intensity dialog. It shows the display parameters setting. You can change the grid and waveform color intensity in this dialog. USER’S MANUAL...
  • Page 68 Hantek 6052BE/6082BE 3.8.4 Waveform Color Click “Display->Waveform Color” in main menu. The following figure shows the waveform color dialog. It shows the waveform color setting. If you click the color rectangle, the color dialog will be showed. The waveform color was changed:...
  • Page 69: Zoom In/Out And Drag Waveforms

    Hantek 6052BE/6082BE 3.9 Zoom In/Out and Drag Waveforms The software will stop updating waveform after the user clicked “Stop” button, The user can change the waveform display by adjusting the scale and position. When you change the scale, the waveform display will increase or decrease in size. When you change the position, the waveform will move up, down, right, or left.
  • Page 70 Hantek 6052BE/6082BE 3.10 Interpolation At the time 40ns/div or faster, user can use the different interpolation mode to get waveforms of different smoothness. The Step Interpolation: The Linear Interpolation: USER’S MANUAL...
  • Page 71 Hantek 6052BE/6082BE The Sin(x)/x Interpolation: Caution: The default interpolation mode is Sin(x)/x. USER’S MANUAL...
  • Page 72: Acquisition Modes

    Hantek 6052BE/6082BE 3.11 Acquisition Modes Acquisition When you acquire a signal, the oscilloscope converts it into a digital form and displays a waveform. The acquisition mode defines how the signal is digitized and the time base setting affects the time span and level of detail in the acquisition.
  • Page 73 Hantek 6052BE/6082BE 1. Click “print” in “File” menu to set the current waveform. 2. Click the “Print Preview” in “File” menu to get into the Preview windows. Click the “Print Preview” in “File” menu to get into the Preview window.
  • Page 74: Chapter 4 Application Example

    Hantek 6052BE/6082BE Chapter 4 Application Example Sample Measurement Pass/Fail Test Capturing a Single-Shot Signal The Application of the X-Y Taking Cursor Measurement USER’S MANUAL...
  • Page 75: Simple Measurement

    Hantek 6052BE/6082BE 4.1 Simple Measurement To acquire and display a signal, please do the steps as follows: Connect signal to CH1 by using probe Click the button on toolbar or “Acquire” -> “Auto Setup” on menu. To measure the frequency and “Vpp”, you can do these steps as follows: 1.
  • Page 76: Pass/Fail Test

    Hantek 6052BE/6082BE 4.2 Pass/Fail Test The Pass/Fail function monitors changes of signals and outputs pass or fail signals by comparing the input signal with the pre-created mask. Control Setting Source: Select the Pass/Fail channel. Output: Select the Pass/Fail output condition.
  • Page 77 Hantek 6052BE/6082BE Vertical: Set the vertical limit range. Horizontal: Set the horizontal limit range. “Create” button: Click this button to create Pass/Fail area according to the mask. “Save” button: Click this button to save the setups to file. “Load” button: Click this button to load the saved setups file.
  • Page 78: Capturing A Single-Shot Signal

    Hantek 6052BE/6082BE Caution: Pass/Fail function unavailable in X-Y mode and Roll mode. 4.3 Capturing a Single-Shot Signal To capture a single event, it needs to gather some pre-test knowledge of the signal in order to set up the trigger level and slope correctly. For example, if the event is derived from 3.3V COMS logic, a trigger level of 1.2 or higher Volts should work on a rising edge.
  • Page 79: The Application Of The X-Y Operation

    Hantek 6052BE/6082BE 4.4 The Application of the X-Y Operation X-Y Plot acts to analyze correlation of data of two channels. Lissajous diagram is displayed in the screen when you use X-Y Plot, which enables to compare frequencies, amplitudes and phases of counterpart waveform against the reference waveform. This makes it possible to compare and analyze frequency, amplitude and phase between input and output.
  • Page 80 Hantek 6052BE/6082BE Instruction of the Ellipse Method: Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals. From the formula above: θ = ±arcsine (A/B) or ±arcsine (C/D) θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between I and III quadrant .If the main axis is at II and IV quadrant, θ...
  • Page 81 Hantek 6052BE/6082BE Measure the Frequency and Time: Read the details showing in the status bar. Measure the Amplitude of the First Waveform Peak of the Waveform Do these steps: 1. Click “Cursor”-> “Source” select CH1 (select CH2 if you want measure CH2).
  • Page 82 Hantek 6052BE/6082BE Measure the Amplitude: Read the details showing in the status bar. Trace the Amplitude of a fixed position on X-axis in a Waveform Do these steps: 1. Click “Cursor->Source”, select CH1 (select CH2 if you want trace CH2).
  • Page 83 Hantek 6052BE/6082BE Trace the Amplitude: Read the details showing in the status bar. Caution: Click “Cursor->Type” to select “Cross”, you can measure time and amplitude at one time. USER’S MANUAL...
  • Page 84: Chapter 5 Appendix

    Hantek 6052BE/6082BE Chapter 5 Appendix Appendix A: Specifications Appendix B: General Maintenance USER’S MANUAL...
  • Page 85: Appendix A: Specifications

    Hantek 6052BE/6082BE Appendix A: Specifications Specifications Table: Input Real-time sampling: Max. sample rate 150MS/s using one channel(Hantek6052BE) 250MS/s using one channel(Hantek6082BE) Channels 50 MHz analog bandwidth(Hantek6052BE) Bandwidth 80MHz analog bandwidth(Hantek6082BE) Vertical resolution 8 bits/channel Gain range 10mV ~5V/div @ x1 probe(10mV, 20mV, 50mV, 100mV, 200mV,...
  • Page 86 Hantek 6052BE/6082BE Settabillity 0.02 div increments Math Vp-p, Vmax, Vmin, Vtop, Vbase, Vmid, Vrms, Vamp, Vmean, Vcyc mean, Positive overshoot, Negative overshoot; Measurements Period, Frequency, Rise time, Fall time, Positive duty cycle, Negative duty cycle, Positive Width, Negative Width Math...
  • Page 87: Appendix B: General Maintenance

    Hantek 6052BE/6082BE Appendix B: General Maintenance General Care Do not store or leave the oscilloscope where the device will be exposed to direct sunlight for long periods of time. Caution To avoid damages to the device or probes, do not expose them to sprays, liquids or solvents.