Rotor
Open the hand lever to provide access to the collet screw. Using a flat blade screwdriver loosen
the collet screw and pull out the rotor assembly, revealing the drive shaft
Fit the collet onto the drive shaft rotating the collet to ensure that the drive dog is correctly located
Push on the rotor, fit and tighten the screw to a torque of 3Nm to prevent the collet from slipping
during operation and close the hand lever
Test the conductivity of the earth bonding between a metal point on
the rotor assembly and the drive casework. The value must be <1MΩ