6. Checklists / Visual inspection

A comment can be added to the results of the visual inspec-
tion and be documented.
Checklists can be configured using the professional software PC200P. Up to 4 different
checklists each with up to 20 items can be stored, edited and documented on the device.

7. Pressure measurements

7.1 Selection of pressure measurements

The following functions can be selected:
High Pressure
Medium Pressure
Micro Pressure
Pump (150 hPa (mbar)) = medium pressure measurement with pump function
High pressure ext. (MPa) = high pressure measurement with external sensor
Differential pressure measurements with the integrated pressure sensor (150 hPa (mbar))
are conducted in the "medium pressure" function.
9
 
 
= pressure measurements up
to 1500 hPa (mbar)
= pressure measurements up
to 150 hPa (mbar)
= pressure measurements up to 100 Pascal
up to 2.5 MPa (25 bar)