Control Box - Mackworth SA180E Service Manual

Stand aid
Table of Contents

Advertisement

Step 4 – If the Plug has not already been removed from the wall socket, remove the plug
from its fixing to the Mast.
Refitting / Replacement
Step 5 - Refitting is a reversal of the removal process noting the following points:
A) Ensure that the snap lead is refitted in the correct orientation in the Control Box
port, and that the snap lead is securely fastened to its latch and will not easily
displace.
9.3 – Control Box
In this section it will explain the correct procedure on removing and reinstalling the Control
Box for servicing procedures or replacement.
Removal
Step 1 – Remove Battery from the Stand Aid. (Refer to section 9.1)
Step 2 – Remove all the Cabling/leads from the Control Box. (Refer to section 9.2)
Step 3 – Use a 3mm Allen Key to remove the Screw fixing the Control Box to the Bracket
Step 4 – Lift the Control Box from its positioning on the Bracket to remove.
Refitting / Replacement
Step 5 - Refitting is a reversal of the removal process noting the following points:
A) Ensure that the Control Box slides into position correctly and vertically straight onto
the Bracket for correct fitting.
995064 – Revision A
Page 18 of 36

Advertisement

Table of Contents
loading

Table of Contents