Products Solutions Services KA01267D/06/EN/02.18 71409635 2018-08-01 Brief Operating Instructions Proline 200 FOUNDATION Fieldbus Transmitter with Coriolis sensor These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Transmitter Brief Operating Instructions Contain information about the transmitter.
Page 2
Proline 200 FOUNDATION Fieldbus Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
Page 3
These Brief Operating Instructions are the Transmitter Brief Operating Instructions. The "Sensor Brief Operating Instructions" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer...
Proline 200 FOUNDATION Fieldbus Document information Document information Symbols used 1.1.1 Safety symbols Symbol Meaning DANGER! DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in WARNING serious or fatal injury.
Page 6
Document information Proline 200 FOUNDATION Fieldbus Symbol Meaning Protective Earth (PE) A terminal which must be connected to ground prior to establishing any other connections. The ground terminals are situated inside and outside the device: • Inner ground terminal: Connects the protectiv earth to the mains supply.
Proline 200 FOUNDATION Fieldbus Basic safety instructions Basic safety instructions Requirements for the personnel The personnel must fulfill the following requirements for its tasks: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task. ‣...
Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device. IT security Our warranty is valid only if the device is installed and used as described in the Operating Instructions.
Installation Proline 200 FOUNDATION Fieldbus max. 350° 8 mm 8 mm A0032242 Release the fixing screw. Turn the housing to the desired position. Firmly tighten the securing screw. Turning the display module The display module can be turned to optimize display readability and operability.
Proline 200 FOUNDATION Fieldbus Installation Transmitter post-installation check The post-installation check must always be performed after the following tasks: • Turning the transmitter housing • Turning the display module Is the device undamaged (visual inspection)? Are the securing screw and securing clamp tightened securely? ...
Electrical connection Proline 200 FOUNDATION Fieldbus Electrical connection Connection conditions 5.1.1 Required tools • For cable entries: Use corresponding tools • For securing clamp: Allen key 3 mm • Wire stripper • When using stranded cables: Crimper for wire end ferrule •...
Page 13
Proline 200 FOUNDATION Fieldbus Electrical connection 5.1.3 Terminal assignment Transmitter Connection version for FOUNDATION Fieldbus, pulse/frequency/switch output A0013570 A0018161 Maximum number of terminals Maximum number of terminals for order code for "Accessory mounted", option NA "Overvoltage protection" Output 1: FOUNDATION Fieldbus...
Page 14
Electrical connection Proline 200 FOUNDATION Fieldbus For reasons concerning explosion protection, it is recommended that grounding be dispensed with. To comply with both requirements, there are basically three different types of shielding in the fieldbus system: • Shielding at both ends •...
Page 15
Proline 200 FOUNDATION Fieldbus Electrical connection A0028768 1 Connection example for FOUNDATION Fieldbus Control system (e.g. PLC) Power conditioner (FOUNDATION Fieldbus) Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable specifications...
Electrical connection Proline 200 FOUNDATION Fieldbus 5.1.7 Preparing the measuring device NOTICE Insufficient sealing of the housing! Operational reliability of the measuring device could be compromised. ‣ Use suitable cable glands corresponding to the degree of protection. Remove dummy plug if present.
Page 17
Proline 200 FOUNDATION Fieldbus Electrical connection Connection via terminals 3 mm 20 mm 10 (0.4) mm (in) A0032239 Loosen the securing clamp of the connection compartment cover. Unscrew the connection compartment cover. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Electrical connection Proline 200 FOUNDATION Fieldbus Removing a cable 3 (0.12) A0032240 ‣ To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between the two terminal holes while simultaneously pulling the cable end out of the terminal.
Proline 200 FOUNDATION Fieldbus Electrical connection To ensure that moisture does not enter the cable entry: Route the cable so that it loops down before the cable entry ("water trap"). A0029278 Insert dummy plugs into unused cable entries. Post-connection check Are cables or the device undamaged (visual inspection)? ...
Operation options Proline 200 FOUNDATION Fieldbus Operation options Overview of operation options A0032226 Local operation via display module Computer with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370 Field Communicator 475 Control system (e.g. PLC)
Proline 200 FOUNDATION Fieldbus Operation options Structure and function of the operating menu 6.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Operation task-oriented Setup Diagnostics Operating menu for experts Expert function-oriented A0014058-EN 2 Schematic structure of the operating menu 6.2.2...
Operation options Proline 200 FOUNDATION Fieldbus Access to the operating menu via the local display X X X X X X 19.184 mA 12.5 X X X X X X Language 20.50 Ã English Deutsch Español Français Xxxxxxx A B C D E F G H I J K M N O P Q R S T U V W +0.000 Xx...
Page 23
Proline 200 FOUNDATION Fieldbus Operation options 6.3.1 Operational display Explanatory symbols for the measured value Status area • Depends on the device version, e.g.: The following symbols appear in the status area of the – : Volume flow operational display at the top right: –...
Page 24
Operation options Proline 200 FOUNDATION Fieldbus Numeric editor Confirms selection. Moves the input position one position to the left. Exits the input without applying the changes. Inserts decimal separator at the input position. Inserts minus sign at the input position.
Proline 200 FOUNDATION Fieldbus System integration Keys and meaning Reduces the contrast (brighter setting). Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). Minus/Plus/Enter key combination (press the keys simultaneously) For operational display:Enables or disables the keypad lock.
Commissioning Proline 200 FOUNDATION Fieldbus • "Post-installation check" checklist→ 11 • "Post-connection check" checklist → 19 Switching on the measuring device ‣ After a successful function check, switch on the measuring device. After a successful startup, the local display switches automatically from the startup display to the operational display.
Proline 200 FOUNDATION Fieldbus Commissioning Configuring the measuring device The Setup menu with its System units submenu and various guided wizards enable fast commissioning of the measuring device. The desired units can be selected in the System units submenu. The wizards systematically guide the user through all the parameters required for configuration, such as parameters for measurement or outputs.
Diagnostic information Proline 200 FOUNDATION Fieldbus Parameter overview with brief description Parameter Description User entry Factory setting Device tag Enter the name for the Max. 32 characters such as Promass 200 measuring point. letters, numbers or special characters (e. g. @, %, /)
Page 29
Proline 200 FOUNDATION Fieldbus Diagnostic information X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2 Diagnostics 3 Supply voltage (ID:203) S801 0d00h02m25s...