User Manual For - CONTROLLER/DATA RECORDER MultiCon CMC-99/141
range* resolution
type
[ mV ]
K
±120
S
±30
J
±120
T
±30
N
±120
R
±30
E
±120
L
±120
B
±30
C
±120
* with consideration of device working in -20'C÷50'C ambient temperature.
Tab. 8.8 Measurement accuracy for the module TC4, TC8, TC12
Tab. 8.9 Thermocouple norm for the module TC4, TC8, TC12
full range
[ μV ]
[
]
°C
4
-200÷1370
1
-50÷1768
4
-210÷1200
1
-200÷400
4
-200÷1300
1
-50÷1768
4
-200÷1000
4
-200÷800
1
250÷1820
4
50÷2290
type
K
S
J
T
N
R
E
L
B
C
full range accuracy
limited range
[
]
[ % ]
°C
±8
±0.51
±7.5
±0.41
±6.3
±0.44
±1.9
±0.31
±12
±0.80
±8.6
±0.47
±4.7
±0.40
±4.4
±0.44
±15
±0.96
±13
±0.51
norm
PN-EN 60584-1:2014
PN-EN 60584-1:2014
PN-EN 60584-1:2014
PN-EN 60584-1:2014
PN-EN 60584-1:2014
PN-EN 60584-1:2014
PN-EN 60584-1:2014
GOST R 8.595:2004
PN-EN 60584-1:2014
PN-EN 60584-1:2014
limited range accuracy
[
]
[
]
°C
°C
-100÷1370
±4
0÷1768
±6
-100÷1200
±3
-100÷400
±1
-100÷1300
±5.9
0÷1768
±5.5
-100÷1000
±2.7
-100÷800
±2.5
480÷1820
±6
50÷1900
±9.2
[ % ]
±0.27
±0.34
±0.23
±0.20
±0.42
±0.31
±0.25
±0.28
±0.45
±0.5