Measuring protocol

Values recorded during installation of Mammomat 300:
Tube P49 Mo, molybdenum anode 0.3
Calculated values
______V x 5 kV/V = _____kV
______V x ____ s x 40 mA/V = ____ mAs
Tube P49 Mo, molybdenum anode 0.15
Calculated values
______V x 5 kV/V = _____kV
______V x ____ s x 40 mA/V = ____ mAs
Keep measuring protocol in technical documentation.
Siemens-Elema AB
Solna, Sweden
SPS-UD
RX B7-120.033.01.10.02
Required values
30 kV +/- 5 %
20 mAs +/- 10 %
Date:______________
Sign:______________
Required values
30 kV +/- 5 %
10 mAs +/- 10 %
Date:______________
Sign:____________