Commissioning; On-Site Commissioning - Pepperl+Fuchs Barcon PPC Operating Instructions Manual

Pressure transmitter
Table of Contents

Advertisement

4

Commissioning

This section contains the following information:
On-site commissioning using keys on the electronic insert
Commissioning and operation using remote communication (Universal
HART Communicator handheld terminal or PACTware)
Locking and unlocking the measuring point
Information on the measuring point
4.1

On-site commissioning

Wire up the Barcon (see chapter 2.5 "Electrical connection")
Connect a multimeter (4 mA ... 20 mA) to the connection terminals
provided.
Ensure that a pressure can be generated within the required measuring
range.
The damping affects the speed with which the output signal and the digital display
react to changes in pressure.
A switch on the electronic insert is used for calibrating the damping:
Switch position off: Damping 0 s
Switch position on: Damping 2 s
Fig. 4.1: Position of the damping switch
Key for calibrating
the zero point
Fig. 4.2: Position of keys for calibrating zero point and measuring span
Zero point calibration is carried out using the key for zero point adjustment (Zero).
Carry out the measuring span adjustment as follows:
Enter exactly the pressure for the lower range value.
Press the Zero key twice.
The acting pressure is adopted as the zero point.
By pressing the Zero key once the value can be called up if a digital display
is connected.
The measuring span is adjusted using the key for calibrating the measuring span
(Span). Carry out the measuring span adjustment as follows:
Enter exactly the pressure for the upper range value.
Press the Span key twice.
The acting pressure for the measuring span is adopted.
By pressing the Span key once the value can be called up if a digital display
is connected.
Damping switch
Damping
Zero
Span
1
2
3
on
off
Key for calibrating
the measuring span
Damping
Zero
Span
1
2
3
on
off
Barcon LHC/PPC (HART)
Commissioning
Contents
Preparatory work
Damping
Zero point adjustment
Adjusting the
measuring span